Technical Data #1851a / 98.06.15
The Enterococci H.L.A.R. Quad Plate, is used to detect enterococci resistance to high level of aminoglycosides and therapeutic synergy with ß-lactam positive antibiotics and glycopeptidic antibiotics. Aminoglycosides and glycopeptides may stop an ordinary enterococci infection but can not stop it if there is an highly resistance to aminoglycosides. Streptomycin, gentamycin and vancomycin are added to a Brain Heart Infusion Agar base.
FORMULA
in grams per liter purified filtered water
| Brain Heard Infusion |
10,0 |
| Casein/Meat Peptone |
17,5 |
| Dextrose |
2,0 |
| Sodium Phosphate, dibasic |
2,5 |
| Sodium Chloride |
5,0 |
| Agar |
15,0 |
|
|
|
| QUAD I: Control X |
|
| QUAD II: Gentamycin |
0,5 |
| QUAD III: Streptomycin |
2,0 |
| QUAD IV: Vancomycin |
0,006 |
|
pH 7,4 +/ñ 0,2 at 25º C |
This approximate formula may be adjusted and/or enriched to obtain best results.
PRECAUTIONS
This medium is for IN VITRO diagnostic use only.
STORAGE
Store prepared media at 2-8º C protected from direct light.
SIGN OF DETERIORATION
Media should not be used if the expiry date has passed. Prepared media should not
be used if there are signs of contamination or deterioration (shrinking, cracking,
evaporation or discoloration).
PROCEDURE
Prior to inoculate, the prepared media should be brought to room temperature.
QUALITY CONTROL
Results after 24 hrs at 35ºC
| Organisms |
ATCC |
QUAD I |
QUAD II |
QUAD III |
QUAD IV |
| Enterococcus faecalis |
29212 |
+ |
- |
- |
- |
| Enterococcus faecalis |
49533 |
+ |
- |
+ |
+ |
| Enterococcus faecalis |
51299 |
+ |
+ |
+ |
+ |
LIMITATIONS OF METHOD
Media should be reincubated for 24 hours more to detect a possible resistance. This
medium is only a part of the identification. Other tests may be required.
REFERENCES
CATALOGUE NUMBERS
| Prepared media | ||
| Plates, 100x15 mm | 10/pkg | 1851 |
| Mono-Plates, 100x15 mm | 10/pkg | 1851M |
| BHI Broth | ||
| Tubes, 2ml, 13X100mm | 10/pkg | 2784 |

| 2331, Dandurand | Phone | Fax | Internet |
| Montreal (Quebec) | (514) 277-2558 | (514) 277-4714 | http://www.quelab.com |